JASCO|Vibrational Circular Dichroism Spectrometer║FVS-6000
Feature :
- Optimized Optics : The FVS-6000 optical system uses a 28 deg interferometer with gold-coated corner cube mirrors for high throughput and to minimize polarization effects with reflective optics to eliminate artifacts caused by birefringence, resulting in very high-quality VCD data.
- Thermal Stability : Multi-zone temperature control with feedback monitoring throughout the optical bench ensures that thermal stability is well maintained for the long-term baseline stability required for long-duration measurements. The key components of the high-intensity ceramic light source), photo-elastic modulator, interferometer, sample compartment, and detector are separately controlled.
- Wide Spectral Range : The FVS-6000 uses a unique, highly sensitive narrow-band MCT detector specially developed for VCD measurement across the spectral range 3,200-850 cm-1. Optional detectors and filters can be added to extend the range to 4,000-750 cm-1, typically for the measurement of OH and NH bands. Detectors are easily exchanged without the need for manual alignment, and optical filters can be selected automatically.
- Narrow Band Mode:The ‘narrow band mode’ allows measurements of small peaks at targeted absorption bands with high sensitivity using optional band-pass filters.
- VCD Auto-Alignment:The auto-alignment function optimizes the VCD optics to reduce linear anisotropy artifacts. No manual alignment is required when the detector and/or optical filters are exchanged.
- Lock-in Detection with DSP:Phase-sensitive lock-in detection uses DSP (Digital Signal Processing) with a sophisticated algorithm optimized for VCD and offers a dramatic improvement in the S/N ratio, particularly compared to more complicated optical designs.
- Purge:The interferometer and optics, sample chamber, and detector housing are all independently purgeable with N2 for high-precision data free of interference from atmospheric gases.
- Long Term stable Measurement:The FVS-6000 MCT detector includes an extra large LN2 dewar for up to 15 hours of operation without refill
|
Model |
FVS-6000 |
|
Measurement Range |
Standard: 3200 ~ 850 cm-1 |
|
Resolution |
16 ~ 0.5 cm-1 |
|
Noise Level (Measurement Conditions) |
less than 8 x 10-6 ΔAbs |
|
Light Source |
High intensity ceramic source, insulated mount |
|
Detector |
Standard: MCT-V (3200 ~ 850 cm-1) |
|
Lock-In Detection |
DSP |
|
PC Communication |
USB2.0 for main unit |
|
Interferometer |
28° Michelson interferometer, with gold coated corner-cube mirrors. mechanical bearing, temperature stabilized |
|
Optical Filter |
Standard: 3200 ~ 2000 cm-1,
2000 ~ 850 cm-1 (6-position automatic switching) |
|
Polarizer |
KRS-5 |
|
Options |
3-position automated sample shuttle |
|
Dimensions (W x D x H) |
590 x 670 x 313 mm |
|
Weight |
65 kg |