JASCO|UV-Visible Spectrophotometer
Feature :
- IQ Accessory and IQ Start: IQ Accessory provides automated accessory recognition with automatic loading of a user defined method. Spectral data files include the instrument and accessory information while designated instrument parameters can be configured and loaded for optimum sample measurements. IQ Start together with the Start Button is used to start a sample method for immediate access when making routine measurements.
- Compact Design: The V-730 is compact with dimensions of only 441 (W) x 468 (D) x 216 (H) mm.
- Simplicity and Ease-of-Use: Both Spectra Manager™ and the iRM provide an intuitive interface allowing first time users to operate the instrument with ease.
- High-Speed Scanning: High-throughput optics and fast response detectors allow the V-730 to scan at speeds up to 8000 nm/min without wavelength tracking errors.
- Sampling Accessories: A wide range of sampling accessories include: autosamplers & sippers, cell holders, flow cells, temperature controllers and software applications.
- Instrument Validation and Regulatory Compliance: The V-730 includes user-selectable validation to USP, EP and JP (or manually editable). With optional IQ/OQ instrument qualification.
- Double-beam spectrophotometer with single monochromator
- Silicon photodiode detectors
- Fixed bandpass 1.0 nm for high resolution measurement
- Dedicated V-730 bio for bio-physical analysis such as kinetics, thermal melting and protein/nucleic acid quant.
|
Model |
V-730 |
|
Optical System |
Single monochromator, Double beam type |
|
Light source |
Halogen lamp, Deuterium lamp |
|
Wavelength range |
190 to 1100 nm |
|
Wavelength accuracy |
+/-0.2 nm (at 656.1 nm) |
|
Wavelength repeatability |
+/-0.1 nm |
|
Spectral bandwidth (SBW) |
1 nm |
|
Stray light |
1 % (198 nm KCL 12 g/L aqueous solution) 0.02 % (220 nm NaI 10 g/L aqueous solution) 0.02 % (340 nm NaNO2 50 g/L aqueous solution) 0.02 % (370 nm NaNO2 50 g/L aqueous solution) |
|
Photometric range |
-3~3 Abs |
|
Photometric accuracy |
+/-0.0015 Abs (0 to 0.5 Abs) +/-0.0025 Abs (0.5 to 1 Abs) +/-0.3 %T Tested |
|
Photometric repeatability |
+/-0.0005 Abs (0 to 0.5 Abs) +/-0.0005 Abs (0.5 to 1 Abs) |
|
Scanning speed |
10-8000 nm/min |
|
RMS noise |
0.00004 Abs |
|
Baseline stability |
0.0004 Abs/hour |
|
Baseline flatness |
+/-0.0005 Abs |
|
Detector |
Silicon photodiode |