JASCO|UV-Visible/NIR Spectrophotometer
Feature :
- IQ Accessory and IQ Start: IQ Accessory provides automated accessory recognition with automatic loading of a user defined method. Spectral data files include the instrument and accessory information while designated instrument parameters can be configured and loaded for optimum sample measurements. IQ Start together with the Start Button is used to start a sample method for immediate access when making routine measurements.
- Wavelength Range from UV to NIR: The V-770 measures the wavelength range of 190 to 2700 nm (option 3200 nm). The unique single monochromator design offers a significantly higher energy throughput resulting in a better S/N ratio over the entire spectral range, even with narrow spectral bandwidth.
- Sampling Accessories: An extensive range of accessories is available for the V-770 UV-Visible/NIR spectrophotometers for virtually any biological, material science or routine QA/QC measurement.
- Options include a wide variety of liquid cell holders, micro and ultra-micro cell holders, flow cell holders and accessories for solid samples. Advanced accessories, including automated cell changers, sippers, integrating spheres, automated and manual absolute reflectance accessories and programmable temperature control systems are available for any of the V-770.
- Instrument Validation and Regulatory Compliance: The V-770 includes user-selectable validation to USP, EP and JP (or manually editable). With optional IQ/OQ.
- Double-beam spectrophotometer with Single monochromator
- Photomultiplier tube detectors and Peltier-cooled PbS.
|
Model |
V-770 |
|
Optical System |
Single monochromator, Double beam type |
|
Light source |
Halogen lamp, Deuterium lamp |
|
Wavelength range |
190 to 2700 nm (Option 3200 nm) |
|
Wavelength accuracy |
+/-0.3 nm (at 656.1 nm) +/-1.5 nm (at 1312.2 nm) |
|
Wavelength repeatability |
+/-0.05 nm (UV-Vis), +/-0.2 nm (NIR) |
|
Spectral bandwidth (SBW) |
UV-Visible: 0.1, 0.2, 0.5, 1, 2, 5, 10 nm L2, L5, L10 nm (low stray light mode) M1, M2 nm (micro cell mode)
NIR: 0.4, 0.8, 1, 2, 4, 8, 20, 40 L8, L20, L40 nm (low stray light mode) M4, M8 nm (micro cell mode) |
|
Stray light |
1 % (198 nm KCL) 0.005 % (220 nm NaI) 0.005 % (340 nm NaNO2) 0.005 % (370 nm NaNO2)
0.04 % (1420 nm: H2O) 0.1 % (1690 nm: CH2Br2) |
|
Photometric range |
UV-Visible: -4~4 Abs NIR: -3~3 Abs |
|
Photometric accuracy |
+/-0.0015 Abs (0 to 0.5 Abs) +/-0.0025 Abs (0.5 to 1 Abs) +/-0.3 %T Tested |
|
Photometric repeatability |
+/-0.0005 Abs (0 to 0.5 Abs) +/-0.0005 Abs (0.5 to 1 Abs) |
|
Scanning speed |
10-4000 nm/min |
|
RMS noise |
0.00003 Abs |
|
Baseline stability |
0.0003 Abs/hour |
|
Baseline flatness |
+/-0.0002 Abs |
|
Detector |
Photomultiplier tube, Peltier cooled PbS |