• JASCO|150 mm Integrating Sphere  for V-700

JASCO|150 mm Integrating Sphere for V-700

Integrating spheres measure either the diffuse transmittance or reflectance of a liquid, solid or powder samples. Typically UV-Visible or UV-Visible/NIR spectrophotometers are used to measure the transmittance of a homogeneous, transparent liquid or solid sample. However, when the light is diffusely scattered by samples such as suspended liquid, opaque solid or powders only a small portion of the light reaches the detector. An integrating sphere captures most of the light from the sample offering a more representative measurement of the sample.
Model : ILV-924、ILN-925、ILN-902i

Feature :

  • Compliant to many international standards that require a larger surface area.
  • Provides a greater measurement for reflectance experiments.
  • Include a Spectralon reference tile and a specular reflectance exclusion sample holder.
  • Provide simple measuring methods for powders, solids and other scattering samples including gem stones.
  • Compliance CIE No.15, ISO 11664-5 ASTM E308, JIS Z 8781-5 etc. colorimetry.
  • Compliance ISO 14782, ASTM D1003, JIS K 7136 haze measurement requirements.
  • Compliance AS/NZS 4399, JIS L1925 BS EN 13758-1, AATCC Test Method 183, GBT18830 UPF Measurement requirements.
  • For JASCO V-700



Item

ILV-924

ILN-925

ILN-902i

Wavelength range

220 to 870 nm

220 to 2200 nm

220 to 1600 nm

Typical Spectral Bandwidth

5 nm

UV-Vis: 5 nm

NIR: 20 nm

UV-Vis: 5 nm

NIR: 10 nm

Sample Size for Cuvette

(Transmittance Measurement)

Light path length 5, 10, 20, 30, 50 mm rectangular cell (sample side)

Light path length 5, 10, 20, 30, 50 mm rectangular cell (sample side)

Light path length 5, 10, 20, 30, 50 mm rectangular cell (sample side)

Sample Size for Solid Sample

(Transmittance Measurement)

Optional SSH-507 Solid Sample Holder need

Minimum 20 (H) x 20 (W) x 0.5 (Thickness) mm

Maximum 70 (H) x 40 (W) x 35 (Thickness) mm

Minimum 20 (H) x 20 (W) x 0.5 (Thickness) mm

Maximum 70 (H) x 40 (W) x 35 (Thickness) mm

Minimum 20 (H) x 20 (W) x 0.5 (Thickness) mm

Maximum 70 (H) x 40 (W) x 35 (Thickness) mm

Sample Size

(Reflectance Measurement)

Minimum 20 (H) x 20 (W) x 0.5 (Thickness) mm

Maximum 100 (H) x 50 (W) x 30 (Thickness) mm

Minimum 20 (H) x 20 (W) x 0.5 (Thickness) mm

Maximum 100 (H) x 50 (W) x 30 (Thickness) mm

Minimum 20 (H) x 20 (W) x 0.5 (Thickness) mm

Maximum 100 (H) x 50 (W) x 30 (Thickness) mm

Powder Sample volume

(PSH-002 Powder Sample Holder need)

φ 16 mm, 0.5 to 6 mm depth

φ 16 mm, 0.5 to 6 mm depth

φ 16 mm, 0.5 to 6 mm depth

Beam

Approx. 9 mm (Height) x 5.5 mm (Width) at the position of the transmittance sample

Approx. 13 mm (Height) x 6 mm (Width) at the position of the reflectance sample

Approx. 9 mm (Height) x 5.5 mm (Width) at the position of the transmittance sample

Approx. 13 mm (Height) x 6 mm (Width) at the position of the reflectance sample

Approx. 9 mm (Height) x 5.5 mm (Width) at the position of the transmittance sample

Approx. 13 mm (Height) x 6 mm (Width) at the position of the reflectance sample

Inside Diameter of Integrating Sphere

150 mm

150 mm

150 mm

Interior of Integrating Sphere

Coated with barium sulfate

Coated with barium sulfate

Coated with barium sulfate

Detector

Photomultiplier tube

Photomultiplier tube, PbS Detector

Photomultiplier tube, InGaAs Detector

Applicable main models

V-750 and V-760

V-770

V-780


Related Products

JASCO|Ambient Accessories for V-700

JASCO|Ambient Accessories for V-700

JASCO|Film Holders for V-700

JASCO|Film Holders for V-700

JASCO|One Drop for V-700

JASCO|One Drop for V-700

JASCO|Water Thermostatted Cell Holder for V-700

JASCO|Water Thermostatted Cell Holder for V-700

JASCO|60 mm Integrating Sphere  for V-700

JASCO|60 mm Integrating Sphere for V-700