JASCO|Fourier Transform Infrared Spectrometers║FTIR-8X
Feature :
- FT/IR-8X spectrometer utilizes a Michelson interferometer with corner-cube mirrors for permanent alignment to prevent light-path deviation, eliminating the need for dynamic alignment.
- The Self-diagnosis function checks the status of the FT/IR-8X at startup. The diagnosis results are automatically recorded, and it is possible to track the temporal change. Therefore, if there is a problem with the data, it can be retraced.
- Built-in NIST traceable polystyrene film for easy validation. The status of the instrument can be checked daily and can be confirmed the reliability of analysis results.
- FT/IR-6X spectrometer controls the interferometer drive using the latest Digital Signal Processing (DSP) technology. his mechanism provides precise moving mirror control and enhances the constant-speed performance of the mirror drive.
- Excellent signal-to-noise ratio (55,000:1) using DSP control and a 24-bit A/D converter.
- The highly accurate 28-degree Michelson interferometer and near-frictionless moving mirror offer class-leading resolution down to 0.07 cm-1.
- The rapid scan provides measurement up to a maximum of 40 Hz.
- Wavenumber range extention option (25000 to 20 cm-1).
- Vacuum option.
Video :
|
Model |
FT/IR-8X |
|
Measurement wavenumber range |
7800 to 350 cm-1 |
|
Measurement wavenumber extended range (optional) |
25000 to 20 cm-1 |
|
Resolution |
0.07, 0.25, 0.5, 1.0, 2, 4, 8, 16 cm-1 |
|
Signal-to-noise ratio |
55000 : 1 |
|
Detector |
Standard:
DLATGS (with peltier temperature control) MCT-N, MCT-M, MCT-W, MCT-PV, Si photodiode (visible, near IR), InSb, InGaAs, DLATGS (PE window), Broad band DLATGS, Si bolometer, DLATGS (for micro measurement)Up to 2 detectors can be mounted inside the main unit, and the external detector unit should be applied if more than 3 detectors are used (PC switching). There are some limited detectors for mounting inside the main unit, and up to 2 detectors can be mounted to external detector unit. |
|
Beam splitter |
Standard:
Ge/KBr Broad band Quartz, Si/CaF2, Broad band KBr, Broad band mylar, Mid-far IR broad band, (exchangeable, optional automatic beam splitter changer is available) |
|
Light source |
Standard:
High-intensity
ceramic source Halogen lamp (PC switching) |
|
Interferometer |
28º Michelson interferometer with corner-cube mirror, sealed structure (KRS-5 window, optional window is also available), auto-alignment mechanism, DSP control, alminum mirror coating |
|
Purging |
Interferometer, sample/detector compartment |
|
Vacuum Instrument |
Sealing |
|
Drive method |
Mechanical bearing, electromagnetic drive |
|
Drive speed |
0.125,
0.25, 0.5, 1, 2, 3, 4, 5, 6, 7, 8 mm/sec |
|
Rapid scan |
40 spectra/sec. (16 cm-1 resolution) |
|
Communication |
USB 2.0 |
|
Dimensions and weight |
600 (W) × 690 (D) × 315 (H) mm, 56 kg |