EYELA|Non-Freon Freeze Dryer
Feature :
- Both FDM-1010 and FDM-2010 can be process up to 1 L of moisture per batch. These unit-based freeze dryers offer flexible system configuration by allowing users to select appropriate manifolds or drying chambers according to the sample vessels.
- Both FDM-1010 and FDM-2010 uses an environmentally friendly natural refrigerant.
- Bright, easy-to-read 3.4 inch LCD display: It features an LCD screen that provides a clear overview of the equipment status, including the trap temperature, vacuum level, and other important operational conditions at a glance.
- Simple operation: In auto mode, the system monitors the cold trap temperature and vacuum level to notify you when the sample vessel is ready to be mounted. Once started, the operation involves only simple steps-mounting the flasks and stopping the process at the end of freeze drying.
- Equipped with timer function: The display of the vacuum pump’s cumulative operating time and freeze-drying duration helps determine the appropriated timing for vacuum pump oil changes and provides a convenient way to measure sublimation time.
- Contamination prevention and nitrogen purge available: The 0.2 ㎛ filter on the air inlet prevents contamination in the event of a vacuum leak. Nitrogen purging is also available to prevent sample oxidation.
- Power failure recovery countermeasures: During power outage recovery, the vacuum pump does not automatically restart. This prevents the sample from scattering due to bumping caused by the sample having thawed due to the power failure.
- Recorder output terminal: for drying data management, the device is equipped with standard output terminals for recording vacuum levels and trap temperatures. The dedicated recorder connection cord is optional.
-
Model FDM-1010 can also be used as a trap: By using the optional cover with nozzle, FDM-1010 can also be used as a trap for centrifugal evaporators (A vacuum pump with a maximum capacity of 533 Pa or less is required.
Video :
|
Model |
FDM-1010 |
FDM-2010 |
|
Cold trap Temperature |
-45 ℃ |
-80 ℃ |
|
Dehumidifying volume |
1 L / batch |
|
|
Setting / display |
Membrane keypad / LCD display |
|
|
Vacuum gauge |
Pirani vacuum gauge (digital display) 0.4 to 533 Pa |
|
|
Safety functions |
Leakage/earth leakage breaker, Refrigerator overload relay, Automatic release upon vacuum leak, Oil change alarm, Fuse for service outlet, Control board self-diagnostic function |
|
|
Vacuum release function |
Automatic vacuum release valve |
|
|
Vacuum pump |
Option: 50 L/min or more |
|
|
Refrigerator / refrigerant |
Air-cooled, output 450W / R290 |
Air-cooled, output 450W x 2 / R290.R170 |
|
Suction port diameter |
17.5 OD (mm) |
|
|
Outer dimensions (mm) |
W450 x D420 x H460 |
W450 (487) x D550 x H945 |
|
Weight |
Approx. 38 kg |
Approx. 83 kg |