HELLMA|Tray Cell
Feature :
- Easy handling – TrayCell always stays in the cell holder
- Excellent reproducibility of the measurement results
- Suitable for almost all standard UV/Vis spectrometers
- For sample volumes from 0.7 μl to 10 μl
- Also suitable for samples with low surface tension
- Different path lengths by means of exchangeable lids
- Long-term precise measurement results without follow-up and maintenance costs
High-tech on the smallest scale - with a patented operating principle
The TrayCell 2.0 comprises a fiber-optic measurement cell and a cap with integrated mirror. One drop (0.7 μl-10 μl) of the sample is pipetted onto the measuring window and then the corresponding cap with the selected path length is placed on top.
The light is guided via prism and optical fiber through the sample, reflected in the mirror and then guided via optical fiber and prism out of the TrayCell 2.0 to the detector.
The precisely defined spacing between the window and the mirror inside the cap ensures a precise and permanently consistent optical path length. The TrayCell 2.0 design makes expensive calibrations and readjustments unnecessary
Representation of the measuring and functional principle of the TrayCell
2.0
|
TYPE |
105.830-UVS |
||
|
Product No. |
105830-A3-V1-40 |
105830-A1-V1-40 |
105830-A2-V1-40 |
|
Path length* |
0.2 mm (factor 50) 1.0 mm (factor 10) |
1.0 mm (factor 10) |
0.2 mm (factor 50) |
|
Window material |
High Performance Quartz Glass |
||
|
Width x depth |
12.5 mm x 12.5 mm |
||
|
Height** |
61.5 mm (center height 8.5 mm) 68 mm (center height 15 mm) 73 mm (center height 20 mm) |
||
|
Volume |
0.7 - 10 μl |
||
|
Max. temperature |
50 °C |
||
|
Center height*** |
8.5 mm, 15 mm or 20 mm |
||
|
Fiber optics |
built in not exchangeable, low solarization 210 nm – 1.100 nm |
||
*
Path length = path length accuracy +/- 0.02 mm, factor = dilution factor
compared to a 10 mm cuvette
**
Height = Total height including cap
***
The center height can be adjusted using the adapters supplied