• JASCO|PDA Detector

JASCO|PDA Detector

The LC-4000 Series includes three PDA detectors. The MD-4017 is a simple replacement for a single wavelength detector, the MD-4015 is the workhorse for UV-Visible detection in the range 200 to 650 nm and the MD-4010 has a wide wavelength range.
The MD-4010 PDA detector has high sensitivity, comparable to a single wavelength detector. It uses a 1024 element photometric diode array, with measurement wavelength range from 190nm to 900nm and resolution of 1 nm. Fast spectral acquisition rate of up to 100 spectra/sec. The temperature controlled SP type semi-micro flow cell is used narrow peaks typically found with fast UHPLC or RHPLC separations. An integrated mercury lamp is used for automatic wavelength calibration and validation.
Model : MD-4010

Feature :

  • Optional Flow Cells

Semi-micro, semi-micro SP (UHPLC), preparative and high pressure for SFC.

  • ChromNAV 2.0 PDA Data System

Includes:  multi-functional analysis with spectral analysis, contour map, 3D, library search for peak identification, peak-purity and multi-wavelength quantitation.

Model

MD-4010

Light source

Deuterium (D2) lamp and Halogen (WI) lamp

Wavelength range

190 to 900 nm

Detector

1024 channels PDA

Slit width

1, 4, and 8 nm, selectable

Noise level

±0.3 × 10-5 AU (under specified conditions)

Drift

±0.5 × 10-3 AU/h (under specified conditions)

Absorbance linearity

2.0 AU or more (under specified conditions)

Data acquisition rate

Maximum 100 points/sec. (full wavelength range)

Flow cell temperature control

Available

PC communication

USB 2.0

Dimensions, weight

300 (W) × 470 (D) × 150 (H) mm, approx. 14.5 kg

Required power

AC 100 to 240 V, 50/60 Hz, maximum 180 VA